Newton Labs Newton Labs
NM200UW measuring head
The NM200UW measurement head

  

Scanner control console
The NM200UW console includes control box, with monitor and keyboard. For transport, the monitor and keyboard are securely stowed within the electronics rack.
Newton Labs
Directional and speed controls for the optional PT200UW Pan-Tilt Arm are
installed on scanner consoles.
Newton Labs

For information on Newton Labs underwater laser scanner products, please contact:
lasersales@newtonlabs.com

Specifications for the NM200UW Nuclear Underwater Laser Scanner

  

Item Control Unit Camera Head
Technical Specifications
System Performance
Item
Control Unit
Measurement Head
Height 14.75 in. (374.65 mm)) 4.250 in. (107,950 mm)
Width 26.75 in. (679.45 mm) 4.625 in. (117,475 mm)
Length 27.50 in. (698.5 mm) 9.375 in. (238,125 mm)
Weight (in air) 84.5 lbs. (38.3 kg) 8 lbs. (4 kg)
Weight (in water) --- 2 lbs. (1 kg) (plus cable weight)
Construction Metal electronics rack suspended on eight shock absorbers within a molded, high-impact, airline-transportable case Machined from solid billet of 6061T6 aluminum stock
Cable LLDPE polyurethane jacket, gel filled - 150 ft. (45.72 m) (other lengths available) ---
Cable weight - 150 ft. (in air) 23.6 lbs. (10.7 kg) ---
Laser power --- 40 mW
Video camera --- High resolution monochrome
LED ring array --- 2,320 lumens
Fittings & retainers --- 300 series stainless steel
Windows --- Fused silica or optical glass
Mounting attachments --- Four grouped 1/4-20 UNC threaded mounting holes on four sides of case (Metric threads available)
Output ports Ethernet, USB, DVI, VGA & HDMI ---
Operating temperature 40° to 110° F (5° to 43° C) 110° F (43.3° C ) in water - 100% duty cycle
Storage temperature 0° to 125° F (-18° to 52° C) 0° to 160° F (- 18° to 71° C)
Power input voltage/current 100 to 240 VAC 50 to 60 cycle Powered by control unit
Data storage Internal solid state & USB stick data ---
Output formats .ply point cloud file ---
Data file size Approximately 100 MB per scan ---
Maximum scanner-to-target distance --- 36.0 in. (900mm)
Minimum scanner-to-target distance --- 6.0 in. (150mm)
Maximum Resolution accuracy
(after processing with 3D software)
--- +/-0.0004 in. (0.01mm)
Scan range --- 6.0 in. to 36 in.
(150 mm to 900 mm)
Watertight depth rating --- 150 ft. (46 m)
Power/Data Umbilical Cable Gel filled gel-filled with LLDPE polyurethane jacket and a rated yield point of 80 lb/ft (11 Kg/M) - 150 ft. (45,7 m) standard (other lengths available)
     

  

System Performance*
System Performance
Depth of Field Distance
(Distance to object)
Field of View
Height ----- Width
Raw Accuracy
(single Point Cloud point, rms)
Approximate CAD Model Accuracy**
6 in. (150 mm)
150 mm 190 mm
+/-.002 in. (0.05 mm)
+/-.0004 in. (0.01 mm)
12 in. (300 mm)
250 mm 330 mm
+/-.004 in. (0.10 mm )
+/-.0008 in. (0.02 mm)
18 in. (450 mm)
350 mm 470 mm
+/-.008 in. (0.20 mm)
+/-.0012 in. (0.04 mm)
24 in. (600 mm)
450 mm 600 mm
+/-.014 in. (0.35 mm)
+/-.0025 in. (0.06 mm)
36 in. (900 mm)
650 mm 880 mm
+/-.028 in. (0.70 mm)
+/-.0060 in. (0.15 mm)
**After 3rd-party 3D software processing
*All NM200UW accuracy is related to the field of view, distance from the object to be measured and can vary by the parameters of the object. Consult Newton for the specific accuracy that can be obtained for your proposed object.
 
Newton Labs